High-Intensity Focused Ultrasound
1
High-Intensity Focused Ultrasound (HIFU)
Contenu
1. Introduction .............................................................................................................................. 2
2. History and medical applications .............................................................................................. 3
2.1 A brief history of HIFU ......................................................................................................... 3
2.2 Tumor ablation applications ................................................................................................ 3
2.2.1 Prostate cancer ......................................................................................................... 4
2.2.2 Uterine fibroids ........................................................................................................ 4
2.2.3 Breast cancer ............................................................................................................ 4
2.2.4 Liver cancer .............................................................................................................. 5
3. Principle .................................................................................................................................... 6
3.1 Thermal effects of HIFU [12] ............................................................................................... 6
3.2 Mechanical effects of HIFU [12] .......................................................................................... 7
3.3 Ultrasound focusing modes ................................................................................................. 7
3.3.1 Mechanical mode ..................................................................................................... 7
3.3.2 Phase-controlled mode ............................................................................................ 8
4. Image-guided HIFU .................................................................................................................. 10
4.1 Ultrasound-guided HIFU .................................................................................................... 10
4.1.1 Treatment planning ................................................................................................ 10
4.1.2 Dose delivery guidance by real-time ultrasound imaging ...................................... 14
4.1.3 Post-treatment assessment by contrast-enhanced ultrasound imaging ................ 15
4.2 MR-guided HIFU (MRgFUS) ............................................................................................... 16
4.2.1 Treatment planning ................................................................................................ 16
4.2.2 MR thermometry ................................................................................................... 22
4.2.3 MR ablation assessment ........................................................................................ 25
4.2.4 MRgHIFU for tumor ablation in mobile organs ...................................................... 26
5. Conclusion ............................................................................................................................... 29
6. References ................................................................................................................................ 30
High-Intensity Focused Ultrasound
2
1. Introduction
High-intensity focused ultrasound (HIFU) therapy is a technology that uses
ultrasound waves to transport energy to target points of body organs, thus increasing
the temperature or bringing about other biological interactions. There will not be any
significant negative biological effects on the intervening tissue as long as that the
ultrasonic energy is appropriately located and focused.
Fig. 1 – High-Intensity Focused Ultrasound (HIFU)
Compared to other hyperthermia methods such as laser, microwave and
radiofrequency, HIFU can treat deep-located tumors in an absolutely non-invasive
manner while sparing peripheral normal tissues.
The most crucial requirement for HIFU therapy is that the thermal ablation must
be precisely delivered to target volume. Because of the complexity and
inhomogeneity of human tissue, it is a challenge to ensure the correct coverage of the
target volume. By integrating HIFU with imaging modalities, it is possible to control
HIFU therapy in order to treat exactly the desired region and minimize damage to
adjacent structures. Currently, MRI and diagnostic ultrasound are being used for
guidance and monitoring of HIFU therapy. Especially, MRI-guided focused
ultrasound surgery (MRgFUS) is quite effective in real-time temperature monitoring,
thus allowing us to conduct closed-loop control of energy deposition.
High-Intensity Focused Ultrasound
3
2. History and medical applications
2.1 A brief history of HIFU
As early as 1927, the biological effects of HIFU were recognized [1]. Then in the
1930s, unfocused ultrasound was applied in physiotherapy [2]. In 1954, Lindstrom [3]
and Fry et al. [4] investigated the possibility of using HIFU for treating neurologic
disorders in humans. Fry and his colleagues were the first to apply HIFU to humans
by producing elevated acoustic intensities in vivo. Then in 1960, they attempted the
use of HIFU for treating Parkinson’s disease [5].
In the 1970s, ultrasound was investigated for treating tumors [6]. Unfortunately,
this strategy didn’t work out mainly due to the lack of a non-invasive temperature
measuring method. In such case, real-time feedback control couldn’t be realized. With
the development of modern technologies, particularly advanced imaging methods
such as MR thermometry, in the 1990s, HIFU gained a rediscovery period [7] [8].
Nowadays, the medical use of HIFU has been permitted in many countries. For
example, MRI-guided treatments of uterine fibroids have been approved in Europe
and Asia, and were granted FDA approval in the US in 2004. Every year, numerous
researches are conducted in the field of HIFU.
2.2 Tumor ablation applications
Not only has HIFU achieved its effective application on tumor ablation, but also
on hemostasis [9], drug delivery [10], and gene expression control [11].
Yet, the major application of HIFU clinically is for the treatment of benign and
malignant solid tumors [12]. HIFU can be applied to numerous kinds of tumors such
as prostate cancer, uterine fibroids, breast cancer, liver cancer and pancreatic cancer.
Here, we discuss several relatively widely used tumor ablations [13].
High-Intensity Focused Ultrasound
4
Fig. 1- Illustration depicts HIFU therapy of intraabdominal tumor [12]
2.2.1 Prostate cancer
Prostate cancer has the longest history of clinical use of HIFU, therefore, many
clinical studies have been performed on it. HIFU for early-stage, localized prostate
cancer has been comparable to surgery in terms of local control, disease-free survival,
and complication rates.
2.2.2 Uterine fibroids
The uterus provides a good target for FUS treatment because it is static and
located close to the abdominal wall. Most clinical studies on uterine leiomyoma have
been done with MR-assisted HIFU devices. HIFU has been shown to be effective in
controlling symptomatic uterine leiomyoma.
2.2.3 Breast cancer
The breast is also a superficial and static organ. However, because of the
difficulties in treating the axillary lymph nodes, the clinical application of HIFU
therapy for breast cancer has been limited.
High-Intensity Focused Ultrasound
5
2.2.4 Liver cancer
The liver, especially the right lobe, is not a suitable organ for the application of
HIFU because of the large respiratory excursions and the sonic shadowing caused by
the ribs. Therefore, most clinical studies have been carried out on palliative
applications rather than for curative purposes.
High-Intensity Focused Ultrasound
6
3. Principle
When ultrasound goes through human tissues, there are two kinds of effects:
thermal effects and mechanical effects [12]. HIFU mainly takes advantage of
ultrasound’s thermal effects to treat tumor, though certain mechanical effects like
cavitation can contribute to the temperature rise of focal region [14]. In the focal
region of the ultrasound transducer, energy is deposited rapidly, causing a rise of the
local temperature, thus generating coagulation necrosis of cancer cells.
Fig. 2 – Acoustic intensity in the focus region [15]
3.1 Thermal effects of HIFU [12]
During HIFU therapy, the temperature raises rapidly to 60℃ or higher in the
High-Intensity Focused Ultrasound
7
tissue. This will cause coagulation necrosis within a few seconds. HIFU generates one
or several focuses in tumor region. In such region, the ultrasonic intensity is much
higher than that outside the focal region. In this way, regions outside the focal will
receive a much lower energy, thus being minimally heated.
3.2 Mechanical effects of HIFU [12]
There are three main mechanical effects: Cavitation, Microstreaming and
Radiation forces.
Cavitation can be defined as the creation or motion of a gas cavity in an acoustic
field. Cavitation occurs due to alternating compression and expansion of tissue as an
ultrasound field propagates through it. There are two forms of cavitation to consider.
The first is stable cavitation, in which a bubble is exposed to a low-pressure acoustic
field, resulting in stable oscillation of the size of the bubble. The other is inertial
cavitation, in which the exposure of the bubble to the acoustic field results in violent
oscillations of the bubble and rapid growth of the bubble during the rarefaction phase,
eventually leading to the violent collapse and destruction of the bubble.
Stable cavitation may lead to a phenomenon called “microstreaming” (rapid
movement of fluid near the bubble due to its oscillating motion). Microstreaming can
produce high shear forces close to the bubble that can disrupt cell membranes and
may play a role in ultrasound-enhanced drug or gene delivery when damage to the cell
membrane is transient.
Radiation forces are developed when a wave is either absorbed or reflected.
Complete reflection produces twice the force that complete absorption does.
3.3 Ultrasound focusing modes
The HIFU system utilizes piezoelectric materials to generate ultrasound. Generally,
it uses an array of piezoelectric cells to form a focus. To adjust the position of the
focus, there are two ways: mechanically or phase-controlled mode.
3.3.1 Mechanical mode
In order to cover the target volume, focus position should be shifted from time to
time. Most HIFU system uses mechanical mode to adjust the position of the acoustic
focus. In such case, the focus position is relatively the same to the piezoelectric cells.
We use a mechanical move system to deliver the focus to positions we desire.
High-Intensity Focused Ultrasound
8
3.3.2 Phase-controlled mode
Phase-controlled mode is also named electronic mode. Every piezoelectric is
controlled by an electronic phase. By adjusting the phases of these cells, we can adjust
the focus’s position. Compared to the mechanical mode, the phase-controlled mode is
more rapid and there are no concerns about mechanical abrasion.
Fig. 3 – Principle of phase-controlled HIFU (A. Spherically-curved transducer, B. Flat
transducer with interchangeable lens, C. Phased-array transducer causing only steering, and D.
Phased-array transducer causing steering and focusing at same time.)
In addition, in phase-controlled mode, we can generate multi-focus pattern, thus
reducing HIFU treatment time under certain circumstances.
High-Intensity Focused Ultrasound
9
Fig. 4 – single-focus and multi-focus mode of phase-controlled HIFU [15]
High-Intensity Focused Ultrasound
10
4. Image-guided HIFU
In order to deliver to the target volume thermal or mechanical energy with high
accuracy, it is indispensable to integrate medical imaging components to the HIFU
system. Several imaging modalities can be utilized for this purpose, thus leading to
the concept of image-guided acoustic therapy. Among these modalities, the most
frequently used are MRI and diagnostic ultrasound. MRI not only provides superior
soft tissue contrast, but also temperature pattern during ablation therapy [16]. While
ultrasound has its advantages mainly in real-time aspect and low cost.
4.1 Ultrasound-guided HIFU
Ultrasonic imaging has the advantage of low cost and excellent real-time
performance. By integrating diagnostic ultrasound with HIFU system, we can 1) get
the spatial information of the tissue and the region of interest; 2) guide ultrasound
energy delivery during treatment; and 3) assess the effectiveness of HIFU therapy
after treatment using contrast enhanced ultrasound imaging.
To illustrate it more clearly, we take prostate cancer HIFU treatment as an
example. Prostate cancer is the most commonly occurring cancer next to skin cancer
and the second leading cause of cancer-related death among men in the U.S [19]. The
number of men diagnosed with prostate cancer is increasing rapidly in recent years. It
is true that HIFU appears to be an attractive and promising therapy in the treatment of
prostate cancer [17].
Focused ultrasound therapy for the treatment of prostate cancer is usually called
transrectal HIFU. The transducer probe is inserted manually into the rectum. If we use
the Sonablate HIFU device, the single piezoelectric crystal alternates between
high-energy power for ablative (3s) and low-energy for ultrasound imaging (6s) which
provides image guidance during treatment [21].
In the case of prostate cancer ablation, there is no significant respiratory-induced
prostate motion [18]. Thus, we don’t take into consideration the motion of prostate
during treatment.
4.1.1 Treatment planning
Before treatment, we should visualize related human organs through diagnostic
ultrasound. It must then be ascertained whether there is a suitable acoustic window
through which the treatment can be delivered, ensuring that the target boundaries can
be clearly identified, and that no sensitive normal tissue structures lie in the beam path.
Then, we should delineate prostate location and relevant treatment zone. To segment
High-Intensity Focused Ultrasound
11
prostate, a semi-automatic method is used. In the meantime, we use Doppler to
determinate treatment zones.
4.1.1.1 Prostate boundary delineation
Prostate volumes and boundaries play an important role in treatment and
follow-up of prostate cancer [20]. For the part of prostate location delineation, there
are a series of segmentation algorithms (mentioned in [20]). Here we cite an accurate
semi-automatic method in the following part [19]. By this method, the mean error
between the computer-generated boundaries and the manual outlining is significantly
less than the manual inter-observer distances.
This method requires prior information of at least two points in the prostate
boundary set by an expert in prostate cancer treatment. Then, from this prior
information, we construct a shape model of prostate using deformable superellipses.
The last step is to use probabilistic theory to final determinate the prostate boundary.
A. Shape model
Deformable superellipses are used to represent prostate shape in 2-D ultrasound
images.
A centered superellipse can be defined in a parametric form by
sin)(
cos)(
{
y
x
ay
ax
and (1)
where the size parameters 0},{ yx aa define the lengths of the semi axes and the
squareness parameter 0 specifies the squareness in the 2-D plane. The implicit
form for a superellipse can be derived from its parametric form as
1)/()/( /2/2 yx ayax (2)
where can be any positive real number if the two terms on the left side of (2) are
first raised to the second power. The inside-outside function which provides a simple
test whether a given point lies inside or outside the superellipse is shown below:
))/()/((),( /2/2 yx ayaxyxf (3)
The superellipse representation is powerful in modeling deformed geometric
shapes through global deformations, such as tapering and bending. A superellipse can
be fully characterized by a vector p = {lx, ly, r, sy, sq, xy, t, b}. The first four pp = {lx,
ly, r, sy} are the pose parameters: translation, rotation, and scaling. The last four ps =
{sq, xy, t, b} define the deformable superellipse shape: squareness sq, aspect ratio xy,
tapering t, and bending b.
High-Intensity Focused Ultrasound
12
B. Model fitting
Superellipse shapes must be fitted to prior manual information in order to
determine the relatively optimal superellipse parameters mentioned above. Here we
illustrate the area-minimizing formulation of the squared algebraic distance method.
N
i
iiyx pyxfaaE
1
2)]1);,(([min (4)
As illustrated in part A, f(xi,yi;p) is the inside-outside function. If its value is below 1,
it means that the point (x,y) is insider the superellipse; while if value above 1, it
indicates that the current point is outside the superellipse. Thus, our goal is to find a
superellipse which minimizes the value of E in equation (4). We take the superellipse
found as a preliminary result shown in Fig. 5.
Fig. 5 – Preliminary result of superellipse fitting
(solid line – manual outlining; dotted contour- deformable superellipses)
C. Prostate segmentation
Using the deformable superellipse as the prior shape model for the prostate, we
segment the prostate boundaries in ultrasound images. The end goal is to find the
optimal parameter vector p = {lx, ly, r, sy, sq, xy, t, b} that best describes the prostate
in a given unsegmented image. The search can be formulated as a maximum a
posterior criterion using the Bayes rule that allows seamless integration of prior shape
knowledge.
We model the shape prior using a multivariate Gaussian distribution Pr(Ps) and the
pose prior using a uniform distribution U(pp). We also assume that shape is
independent from pose based on our modeling study. The edge strength E of an image
can be used as the likelihood. By using the Bayes rule, the a posterior probability
density of the deformed boundary given the input edge strength can be expressed as
)Pr(
)|Pr()Pr()(
)Pr(
)|Pr()Pr(
)|Pr(
E
pEppU
E
pEp
Ep
sp
(5)
High-Intensity Focused Ultrasound
13
Pr(E) is the prior probability of the edge strength which can be derived from the
image data, thus Pr(E) is equal for all p. U(pp) is constant for all p. In such case, we
take the logarithm form of Pr(p|E) and find that it is only related to Pr(ps) and Pr(E|p).
As illustrated in [19], the problem now is to maximize the following function
nj
jjj npynpxEmpf )),(),,(()]2/()([
22 (6)
mj is the mean value of pj (mj is from the preliminary result), and pj is the jth element
of the shape parameter vector ps. The variance for each of the parameters is calculated
from the training set. (x(p,n), y(p,n)) is the pixel location of the nth pixel in the
deformable superellipse contour. is the weighting parameter generally associated
with the image noise model.
In this way, we can get the corresponding parameter vector p of the maximum
value of f. In other words, we get the final optimal superellipse. Fig. 6 is two
examples of the comparison between manual outlin