Designation: D 6247 – 98
Standard Test Method for
Analysis of Elemental Content in Polyolefins By X-Ray
Fluorescence Spectrometry1
This standard is issued under the fixed designation D 6247; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers a general procedure for the
determination of elemental content in polyolefins by X-ray
fluorescence spectrometry, in concentration levels typical of
those contributed by additives and reactor processes.
NOTE 1—Specific methods and capabilities of users may vary with
differences in interelement effects and sensitivities, instrumentation and
applications software, and practices between laboratories. Development
and use of test procedures to measure particular elements, concentration
ranges or matrices is the responsibility of individual users.
NOTE 2—One general method is outlined herein; alternative analytical
practices can be followed, and are attached in notes, where appropriate.
1.2 The values stated in SI units are to be regarded as the
standard. The inch-pound units given in parentheses are for
information only.
1.3 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use. Specific precau-
tionary statements are given in Section 10.
NOTE 3—There is no similar or equivalent ISO standard.
2. Referenced Documents
2.1 ASTM Standards:
C 1118 Guide for Selecting Components for Wavelength-
Dispersive X-Ray Fluorescence (XRF) Systems2
D 883 Terminology Relating to Plastics3
D 4703 Practice for Compression-Molding Thermoplastic
Materials into Test Specimens, Plaques or Sheets4
E 135 Terminology Relating to Analytical Chemistry for
Metals, Ores, and Related Materials5
E 691 Practice for Conducting an Interlaboratory Study to
Determine the Precision of a Test Method6
E 1361 Guide for Correction of Interelement Effects in
X-Ray Spectrometric Analysis7
E 1621 Guide for X-Ray Emission Spectrometric Analysis7
3. Terminology
3.1 Definitions:
3.1.1 Definitions of terms applying to XRF and plastics
appear in Terminology E 135 and Terminology D 883, respec-
tively.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 element—used in this context, refers to any chemical
element that can be determined by XRF; and is often used
synonymously with the term metal.
3.2.2 infinite thickness—or critical thickness: the thickness
of specimen which, if increased, yields no increase in intensity
of secondary x-rays, due to absorption by the polymer matrix.
This thickness varies with secondary x-ray energy, or wave-
length.
3.2.3 polyolefin—used in this context, refers to PE and PP
thermoplastics.
4. Summary of Test Method
4.1 The test specimen is compression molded or injection
molded into a plaque having a clean, uniform surface.
4.2 The plaque is irradiated in the WDXRF with a beam of
primary X-rays, that causes each element to fluoresce at
specific wavelengths (lines). Choice of appropriate lines and
spectrometer test conditions can vary according to each ele-
ment, and with factors such as detector response, concentration
range, and other elements present in the sample matrix.
4.3 These secondary X-rays are dispersed with diffraction
crystals of appropriate spacing, and measured by appropriate
detectors configured at angles specific to lines of interest.
Additional considerations appear in Guides C 1118 and
E 1621.
4.4 Analyte concentration is determined by relation/
comparison of line intensity with a calibration curve.
1 This test method is under the jurisdiction of ASTM Committee D-20 on Plastics
and is the direct responsibility of Subcommittee D20.70 on Analytical Methods.
Current edition approved April 10, 1998. Published February 1999.
2 Annual Book of ASTM Standards, Vol 12.01.
3 Annual Book of ASTM Standards, Vol 08.01.
4 Annual Book of ASTM Standards, Vol 08.03.
5 Annual Book of ASTM Standards, Vol 03.05.
6 Annual Book of ASTM Standards, Vol 14.02.
7 Annual Book of ASTM Standards, Vol 03.06.
1
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
NOTE 4—An alternative method utilizes a fundamental parameters type
calibration.
5. Significance and Use
5.1 X-ray fluorescence spectrometry is an accurate and
relatively fast method to determine multielement concentra-
tions in polyethylene or polypropylene materials, or both. It
can be used as a quality control measure for post-reactor
studies, for additive levels in formulated resins, and for
finished products.
5.2 Further elaboration appears in Guide E 1621.
6. Interferences
6.1 Potential interferences are explained in Guide E 1621.
7. Apparatus
7.1 Calibration Standards Formulation:
7.1.1 Batch Compounding Equipment, with temperature
regulation capabilities, for melt homogenization of elemental
compounds or additives into polyolefin reference standards.
Equipment can range from small scale torque rheometers
equipped with mixing head, to larger-scale batch mixers.
Nitrogen purge capabilities are recommended.
NOTE 5—An alternative method requires a single-screw or twin-screw
laboratory-scale extruder in place of the melt-fusion head, however, more
material is required for formulation. Dry homogenization techniques
which do not require the use of melt-compounding apparatus have been
used, however, such are not recommended.
7.1.2 Analytical Balance, 0.1-mg sensitivity
7.2 Specimen Preparation:
7.2.1 Thermal Press, for compression-molding of plaques,
and capable of obtaining temperatures, pressures and cooling
rates, as recommended for PE and PP in Practice D 4703 and
in Section 11 of this test method.
7.2.2 Flash Type Mold, picture-frame type, described in
Practice D 4703: stainless-steel chase to mold test plaques 3.2
mm (0.125 in.) thick, uncoated Mylar8 (polyester film) parting
sheets, and smooth, stainless steel backing plates of minimum
2.5 mm thickness.
NOTE 6—Injection molding apparatus have also been employed, in
place of the thermal press and flash mold.
NOTE 7—The recommended thickness of 3.2 mm may vary with user
preference. Other material may be used for parting sheets, provided it does
not adhere to the polymer under such thermal conditions. Some techniques
use polished steel backing plates, with chromed surfaces, and no parting
sheets. The apparatus used should not introduce contamination of ele-
ments of interest, and should yield specimens with a plane surface of
adequate smoothness.
7.3 Excitation Source, Spectrometer, and Measuring
System—Requirements for a wavelength-dispersive XRF are
outlined in Guides C 1118 and E 1621.
8. Reagents and Materials
8.1 P-10 Gas, a mixture of 90 % argon and 10 % methane,
ultra-high purity or equivalent, for use with gas-flow propor-
tional detectors.
8.2 Nitrogen, prepurified grade or equivalent, for purging
the melt fusion chamber.
8.3 Elemental Standards—Compounds or additives, or
both, to be melt homogenized into polymer calibration stan-
dards. Material should be certified as to elemental purity and
should be dried and desiccated prior to use.
9. Reference Materials
9.1 Reference Materials, can be prepared by users, as in
Section 11.
9.2 Other Materials, appear in Guide E 1621.
9.3 Monitor Samples—To account for instrument drift (for
example, electronic variations or long term X-ray source
degradation).
NOTE 8—Monitor samples are selected for their long-term durability,
and should ideally contain elements near to high and low ends of
analytical range(s) of interest. Prefabricated glass discs outside range(s) of
interest are often used. They are measured before measurement of
calibration standards, and are remeasured periodically (for example,
daily). Any drift from original measurements can be accounted for, and
necessary corrections can be applied to maintain calibrations relative to
initial conditions.
10. Safety Precautions
10.1 Occupational Health and Safety Standards for X-Rays,
and ionizing radiation shall be observed. It is also recom-
mended that proper practices be followed, as shown in Guide
E 1621.
NOTE 9—X-rays are dangerous and can cause serious personal injury.
X-ray beams can be very narrow and difficult to detect. Precautions taken
minimize potential radiation exposure include an increase in protective
shielding, an increase of distance, and a decrease of time near any
suspected source of leakage.
10.2 Chemical—Appropriate precautions for chemical haz-
ards shall be observed for any chemicals and materials used in
preparing calibration standards. Consult the suppliers’ Material
Safety Data Sheets for specific hazards and safety practices.
10.3 Pressurized Gas requires safe and proper handling
practices.
10.4 Specimen Preparation—Follow appropriate precau-
tions when using hot equipment for homogenization and
specimen preparation. Consult manufacturer’s recommenda-
tions for specific practices.
11. Preparation of Reference Materials and Test
Specimens
11.1 Specimen Plaques—Consistent preparation of refer-
ence materials and test specimens is essential. Variations in
sample thickness, surface finish and homogeneity can affect
reliability of results.
11.1.1 Reference Materials—A technique that offers consis-
tent elemental dispersion throughout the calibration standard
should be followed. Melt homogenization to ensure uniformity
is recommended; see Annex A1.
NOTE 10—Resins from actual production runs have been used for
calibration standards, after verification by independent analytical methods.
11.1.2 Test Specimens—If elemental dispersion is poor,
homogenize the sample according to Annex A1, Section 2. If
elemental dispersion is acceptable, proceed to 11.2.8 Mylar is a registered trademark of DuPont.
D 6247
2
11.2 Compression Molding—Both calibration standards and
test specimens are pressed into plaques having a suggested
thickness of 3.2 mm (0.125 in.). Each plaque should have a
smooth, plane surface, and no voids. Refer to Practice D 4703.
NOTE 11—Techniques using no polymer-based calibration have been
used.
NOTE 12—Some users employ a laboratory-scale injection molding
technique as an alternative to this suggested compression molding
method.
NOTE 13—A thickness of 3.2 mm is commonly used, however, some
laboratories employ lesser or greater thicknesses (for example, 6 mm, to
more closely approach infinite thickness; or extruded thin films and
sheets). The convenience of making a plaque of one thickness for all
elements, versus infinite thickness may be a factor for user consideration.
NOTE 14—Plaques may be molded in the proper dimensions to fit the
instrument, or may be of a larger size, from which the specimen plaque is
punched out.
11.2.1 Weigh the required mass of plastic and place into the
hollow chase of the picture frame flash mold assembly, which
is sandwiched between Mylar8 parting sheets, and supported
by smooth backing plates.
11.2.2 Place the assembly into the press, which has been
preheated to 175°C for PE and 200°C for PP. Heat the sample
under appropriate pressure and time settings to obtain a
suitable specimen plaque.
11.3 Cooling Rate is chosen to yield a uniform, smooth
plaque. User preferences vary.
11.3.1 Programmed Cooling—After a pre-programmed
dwell time at high pressure, initiate cooling at a recommended
rate of 15° 6 2°C per minute, or:
11.3.2 Quick Cooling—Remove the hot assembly and pres-
surize in a water-cooled press, or equivalent, and allow it to
cool to ambient temperature.
11.4 It is recommended to clean specimen surfaces with
isopropyl alcohol (or methyl alcohol), immediately prior to
measurement. Operator experience may show this need to vary
with the element or additive to be measured.
11.4.1 Care should be taken to handle only the sides, not the
surface of the specimen following molding or cleaning so that
oils and salts from the skin do not contaminate the specimen.
Disposable cotton gloves may be worn when handling speci-
mens to prevent inadvertent contamination.
12. Preparation of Apparatus
12.1 A description of considerations is included in Guide
E 1621.
NOTE 15—Measurement of Ka lines is recommended, where possible.
NOTE 16—A reduction of X-ray tube power used (for example -10 %
current) can result in an overall increase in tube life.
13. Calibration and Standardization
13.1 Considerations are covered in Guide E 1621.
NOTE 17—Background subtraction is a suggested correction method for
polyolefins. A linear calibration curve was suitable for the elemental
concentrations listed in the annex. Higher concentrations may require
mathematical corrections covered in Guide E 1361.
NOTE 18—One alternative calibration method, using fundamental pa-
rameters software, required measurements of only one polymer blank and
one standard per element (for example, 0.5 % analyte element in mineral
oil).
14. Procedure
14.1 Refer to Guide E 1621.
NOTE 19—The effects of poor surface finish on test accuracy can
become more adverse at longer wavelengths (lower atomic numbers) and
at lower analyte concentrations. The use of a sample spinner is recom-
mended.
14.2 It is recommended to measure monitor sample(s) daily,
for elements of interest, and to apply any required drift
corrections to sample measurements.
14.3 Participating laboratories preferred a fixed counting
time over a fixed number of counts.
15. Interpretation of Results
15.1 Refer to Guide E 1621.
16. Precision and Bias
16.1 Instrument Repeatability—Repeatability of one instru-
ment in one laboratory measuring one single specimen of an
experimental formulation 30 times is shown in Table 1.
16.2 Round Robin Study:
16.2.1 Precision and bias statements are currently under
evaluation. Data were obtained in a round robin study con-
ducted in 1992, in accordance with Practice E 691, involving
two groups of commercial grade materials (one group of
polyethylene, one group of polypropylene), with each group
containing three materials (that is a total of six materials). The
grades were selected on the basis of their relative elemental
content levels, from low (< 10 ppm), to medium (< 200 ppm),
to high (> 200 and< 5000 ppm); as contributed by catalyst
residues or additives. Each grade was relatively free of
interferences and interelemental effects. Reference values for
the elements selected for testing in each of the round robin
samples were determined by neutron activation analysis.
16.2.2 To minimize variations in elemental dispersion, each
sample material was homogenized by triple extrusion passes.
Polyethylene and polypropylene sample groups were prepared
by each of two laboratories respectively, and distributed to
each round robin participant under cover of these respective
labs. Twelve laboratories participated in this study. Each round
robin participant had the option to analyze either one, or both
groups of materials. The polyethylene sample group was tested
for a total of nine elements; the polypropylene sample group
was tested for a total of eight elements. Each sample was tested
for a specified set of elements selected from within its
respective group.
16.2.3 Each sample was analyzed in duplicate on each of
three separate days (that is a total of six determinations per
sample, of each specified element). Each laboratory conducted
measurements only on elements it analyzed for on a routine
basis. All testing was conducted on WDXRF instruments.
TABLE 1 Instrument Repeatability
Element Mean Concentration
(ppm)
Standard Deviation
P 18.9 0.20
Si 1538 3.4
Zn 58.1 0.22
D 6247
3
16.3 The following explanations of r and R (16.4-16.4.3 are
intended to present a meaningful way of considering the
approximate precision of this test method, as applicable to each
respective element reported, and at the concentration levels
outlined in Table 2. The data under evaluation should not be
applied rigorously to acceptance or rejection of a material, as
those data are specific to the round robin, and may not be
representative of other lots, conditions, materials, or laborato-
ries. Users of this test method should apply the principles
outlined in Practice E 691 to generate data specific to their
laboratory and materials, or between specific laboratories. The
principles of 16.4-16.4.3 would then be valid for such data.
16.4 Concept of r and R—If Sr and SR have been calculated
from a large enough body of data, and for test results that were
averages from testing six specimens, the following applies:
16.4.1 Repeatability Limit, r (Comparing Two Test Results
for the Same Material, Obtained by the Same Operator Using
the Same Equipment on the Same Day)—The two test results
should be judged not equivalent if they differ by more than the
r value for that material.
16.4.2 Reproducibility Limit, R (Comparing Two Test Re-
sults for the Same Material, Obtained by Different Operators in
Different Laboratories)—The two test results should be judged
not equivalent if they differ by more than R value for that
material.
16.4.3 Any judgement in accordance with 16.4.1 or 16.4.2
would have an approximate 95 % (0.95) probability of being
correct.
16.5 Bias (Accuracy)—Capabilities of individual laborato-
ries resulted in differing numbers of participants contributing to
measured data, depending upon each particular element, and its
concentration. Table 2 shows referee concentrations, as deter-
mined by neutron activation analysis.
17. Keywords
17.1 element; polyolefin; X-ray fluorescence spectrometry
(XRF)
ANNEX
(Mandatory Information)
A1. PREPARATION AND MELT HOMOGENATION OF SPECIMENS
A1.1 Preparation of Reference Standards—Accurately
weigh appropriate quantities of respective additives/
compounds into the dried, granular polymer.
A1.2 Melt Homogenization is used to minimize local
variations of additive/element concentration. Batch compound-
ing equipment is recommended to ensure accurate mass
balance(s) throughout the mixing process. A torque rheometer
equipped with mixing head, or a larger melt compounder is
suitable.
NOTE A1.1—Although batch compounding is recommended, some
users melt-homogenize via single or multiple passes of the polyolefin
standard through a laboratory-scale extruder.
NOTE A1.2—Dry homogenization requires less apparatus, but may
yield less uniform standards. This technique is not recommended. Ele-
ments or additives, or both, are dissolved in appropriate solvents, in
specified concentrations. Aliquots of appropriate volumes are distributed
evenly into preweighed quantities of additive-free granular resin, in
respective evaporating dishes. These are manually stirred while evaporat-
ing to dryness on a non-flammable heat source such as a steam bath.
A1.2.1 Calibration Standards must be melt homogenized
prior to their use as reference materials for calibration.
A1.2.2 Test Specimens—Melt homogenization is recom-
mended for test specimens.
A1.2.3 Antioxidant Stabilization of sufficient concentration
to prevent thermal degradation of base resins during homog-
enization is required. The antioxidant used should not contain
elements which interfere with elemental lines of interest.
A1.2.4 Blending should be carried out under conditions
suitable to the resin and equipment used. Temperatures 30 to
50°C above resin melt temperatures are recommended. Mixing
rate and time are chosen to ensure adequate uniformity and
minimal degradation of samples. The use of a nitrogen purge is
recommended.
NOTE A1.3—Purge with nitrogen at a rate which provides an inert
nitrogen blanket.
NOTE A1.4—Operator experience may dictate adjustment of blending