7.5Fr球囊穿刺步骤
FOR SHEATHLESS INSERTION FOR SHEATHED INSERTION
PRESSURE MONITORING SET UP
The angle of insertion should be
45˚ or less.
Make a small incision at the exit
of the guidewire.
Insert the vessel dilator over the
guidewire, tapered end first.
Push then twist to...
FOR SHEATHLESS INSERTION FOR SHEATHED INSERTION
PRESSURE MONITORING SET UP
The angle of insertion should be
45˚ or less.
Make a small incision at the exit
of the guidewire.
Insert the vessel dilator over the
guidewire, tapered end first.
Push then twist to firmly attach
the one-way valve to the male
luer fitting of the IAB catheter.
Using the provided 30cc syringe
apply a 30cc vacuum.
Twist and remove the syringe, keeping
the one-way valve attached to the male
luer fitting of the IAB catheter.
Carefully spread the
subcutaneous tissue
to facilitate sheathless
insertion.
Remove the IAB catheter
from the T-handle by
pulling STRAIGHT out
to avoid damaging it.
Advance the IAB catheter
over the guidewire into
the artery in short strokes
until correct placement
is achieved.
Advance the sheath seal as
close to the insertion site as
possible and secure the IAB
catheter to the patient’s leg.
Insert the introducer
dilator into the sheath
hub and secure.
Advance the sheath
over the guidewire into
the artery using a rotary
motion. Withdraw the
introducer dilator leaving
the sheath in place.
Remove the IAB catheter
from the T-handle by
pulling STRAIGHT out
to avoid damaging it.
Advance the IAB catheter
in short strokes through
the sheath into the artery
until correct placement is
achieved. Push the sheath
seal into the hub of the
sheath and secure the IAB
catheter to the patient’s leg.
Remove the guidewire and
aspirate 3cc of blood from the
inner lumen.
Perform a manual flush of the
inner lumen with 3-5cc of flush
solution.
Attach a standard arterial fluid
pressure monitoring apparatus.
Twist and pull to remove the
one-way valve from the IAB
catheter.
Connect the IAB catheter’s male
luer fitting to the female luer
fitting of the catheter extender.
Connect the male luer fitting of the
catheter extender to the safety disk
of the Intra-aortic Balloon Pump.
INSERTION OF THE INTRA-AORTIC BALLOON CATHETER
ABBREVIATED GUIDE FOR LINEAR 7.5Fr. AND MEGA 8Fr.
IAB CATHETER INSERTION
This guide is a reference for personnel trained in using MAQUET
IAB Catheters. Do not use this guide until after thoroughly reading
the Instructions For Use provided.
Note: This information is to be used as a guidance only. Clinical
information and patient factors such as torso length should be
considered when selecting the appropriate balloon size.
1
PREPARING THE IAB CATHETER
CONNECTION TO IABP
INTRA-AORTIC BALLOON
SIZING GUIDE
2 3 4 5 6
7 8 9 10 107 8 9
11 12 13 14 15 16
6'
5'
4'
Approx. ht: < 5'0"
(< 152cm)
Approx. ht: 5'0" to 5'4"
(152cm - 162cm)
Approx. ht: 5'4" to 6'0"
(162cm - 183cm)
Approx. ht: ≥ 5'4"
(≥ 162cm)
25cc 34cc 40cc 50cc
® MAQUET Registered Trademark of MAQUET GmbH & Co. KG · Reg. QUG: MCV-PT-80000308-EN-01 · PN: 0002-08-1002 R0 · Copyright 2010 MAQUET Cardiovascular LLC or its affiliates · Subject to modifications due to technical development. 3.10.
Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician. Refer to Instructions for Use for current indications, warnings, contraindications, and precautions.
DATASCOPE IS NOW MAQUET CARDIOVASCULAR
Datascope Cardiac Assist
15 Law Drive
Fairfield, NJ 07004
Phone: +1 973 244 6100
Please visit our website:
www.datascope.com
www.maquet.com
本文档为【7.5Fr球囊穿刺步骤】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。